Test of solution of fluorescein

Goal

Check the absorbance and fluorescence of a solution of fluorescein in order to find the lowest concentration required to have a high fluorescence.

Results

Spectro 1

Concentration [µg/L]ERGBUV
140.250.0080.2981.74313977
70.470.0120.2131.64235406
1.40.680.0050.0881.03666273
0.70.890.0300.0530.63055499
0.351.11-0.0010.0300.34037857
0.141.410.1130.1950.16418380
0.071.630.0070.0140.07310000

Spectro 2

Concentration [µg/L]ERGBUV
140.240.0050.2941.73815022
70.480.0110.2081.64135132
1.40.780.0030.0851.03869277
0.71.020.0270.0480.63156721
0.351.26-0.0020.0250.33938510
0.141.510.1090.1720.16618355
0.071.770.0070.0180.08110121
0.0142.030.003-0.0060.0142019
0.00142.27-0.006-0.0140.003110
06.710.0040.0020.00016
0.00146.96-0.001-0.0130.007110
07.180.0110.0120.0080
0.00147.39-0.006-0.0150.002108
07.630.0090.0100.00711
0.00147.86-0.008-0.0150.001105

All data

Conclusions

We are able to detect 1.4µg per liter of fluorescein.