Starting a new bioreactor

When you have a new bioreactor you should start by configuring it. You may connect using a terminal through the serial port.

You may as well use the Visual Studio Code (with PlatformIO extension) application to directly connect to the bioreactor through serial (press (Mac) ++P or (PC/Linux) Ctrl++P and search for serial).

The menu is accessible using the h instruction.

important

If you can't open the serial monitor in VSCode with PlatformIO extension, please make sure that you have opened the project.

caution

You should not forget to add a CR + LF after each instruction, this is automatically in VScode Serial Monitor but you need to consider it if you don't have an answer in the serial monitor.

Initializing parameters

Connect via the terminal and reset all the parameters to the default parameters:

ur1234

It is also important to set a unique qualifier for the bioreactor. You should fill the list with a free code.

Setup the weight

There is now a special menu for the weight calibration w.

You should do the following step in order to callibrate the weight:

  1. Empty bioreactor : we
  2. Empty bioreactor + 1kg : wk
  3. Bioreactor filled at high level : wh

You may now check the reproducibility and the weight in g of any object using wt

It is important to note that an error (stored in PARAM_ERROR) will be generated if the weight is either 20% under the minimal value or 20% over the maximal value and this should stop all the functions.

The main parameters

In order to activate the bioreactor you need to enable functions. Currently there are 3 functions:

  • Agitation (value 1)
  • Food control (value 2)
  • Temperature control (value 3)

You may enable the functionnablities by setting the parameter AZ. To activate all the functionnalities enter AZ63.